Wiring as cocked up and then misdescribed by Sandowne is in fact this...
N.B. All 25-way cables are straight through. Older Camden cables have different pinouts, so use RNC's map (or similar).
Note: black/grey is different from black/dirty white and black/cyan, though they are very similar!
(Black/grey is pin 24, ground; I think black/white is pin 22, unused.)
Device
|
Box terminal#
|
25-way connector
|
Wire colour (Maria's 25-way straight-through cables)
|
Camden pin#
|
Camden panel label
|
Left lever report
|
IN 1
|
9
|
yellow
|
1
|
Left lever report
|
Right lever report
|
IN 2
|
10
|
yellow/black
|
5
|
Right lever report
|
Tray IR detector
|
IN 3
|
11
|
dark green
|
3
|
Tray report
|
Photobeam #1 FRONT
|
IN 4
|
12
|
dark green/white
|
4
|
Spare 5
|
Photobeam #2 MIDDLE
|
IN 5
|
25
|
white
|
6
|
Spare 6
|
Photobeam #3 READ
|
IN 6
|
13
|
cyan (or light green!)
|
18
|
Spare 2
|
Left lever operate
|
OUT 1
|
1
|
brown
|
17
|
Spare 1
|
Right lever operate
|
OUT 2
|
15
|
blue
|
19
|
Spare 3
|
Liquid dipper
|
OUT 3
|
2
|
brown/white
|
7
|
Reinforce
|
Left light
|
OUT 4
|
16
|
blue/white
|
13
|
Left light
|
Centre light
|
OUT 5
|
3
|
red
|
|
|
Right light
|
OUT 6
|
17
|
light blue
|
15
|
Right light
|
Houselight
|
OUT 7
|
4
|
red/white
|
16
|
Houselight
|
Syringe pump
|
OUT 8
|
18
|
light blue/black
|
12
|
Traylight
|
Traylight
|
OUT 9
|
5
|
pink
|
|
|
Clicker
|
OUT 10
|
19
|
purple
|
|
|
|
OUT 11
|
6
|
|
|
|
|
OUT 12
|
20
|
|
|
|
|
OUT 13
|
7
|
|
|
|
|
OUT 14
|
21
|
|
|
|
|
OUT 15
|
8
|
|
|
|
|
OUT 16
|
22
|
|
|
|
|
+28V
|
23
|
grey
|
|
|
|
28V GND
|
24
|
grey/black
|
|
|
|
|
|
|
|
|
Camden pin 20 (spare 4) – not connected
Not used on the Camden panels: spare 4 / loudspeaker / centre
Must connect Camden panel COMMON and OPERATE to ground before the boxes will work.
Sandowne had described RIGHT LIGHT as connected to OUT 5, and had connected it to OUT 5, but they should have used OUT 6. Oops.
|